First-in-man implantation of the retrievable and repositionable VenusA-Plus valve

2018 
BACKGROUND: No retrievable and repositionable second generation transcatheter aortic valve is available in China. Here, we report the first-in-man implantation of the retrievable and repositionable VenusA-Plus valve. METHODS: A 76-year-old patient with symptomatic severe aortic stenosis and high surgical risk (STS 13.8%) was recommended for transcatheter aortic valve replacement (TAVR) by heart valve team. Type 0 bicuspid aortic valve with asymmetric calcification was identified by dual source computed tomography, and the unfavorable anatomies increased the possibility of malposition and paravalvular leakage during TAVR. Therefore, we used the retrievable and repositionable VenusA-Plus valve for the patient. RESULTS: Transfemoral TAVR was performed under local anesthesia with sedation, and a 26mm VenusA-Plus valve was successfully implanted. No transvalvular pressure gradient and trace paravalvular leakage were found. CONCLUSION: The successful first-in-man implantation indicates the retrievable and repositionable VenusA-Plus valve is feasible in complicated TAVR cases such as bicuspid aortic valve.