Photon-counting Computed Tomography of Lanthanide Contrast Agents With a High-Flux 330- μ M-Pitch Cadmium Zinc Telluride Detector in a Table-Top System

2020 
Purpose: We present photon-counting computed tomography (PCCT) imaging of contrast agent triplets similar in atomic number ( Z ) achieved with a high-flux cadmium zinc telluride (CZT) detector. Approach: The table-top PCCT imaging system included a 330 - μ m -pitch CZT detector of size 8 mm × 24 mm 2 capable of using six energy bins. Four 3D-printed 3-cm-diameter phantoms each contained seven 6-mm-diameter vials with water and low and high concentration solutions of various contrast agents. Lanthanum ( Z = 57 ), gadolinium (Gd) ( Z = 64 ), and lutetium ( Z = 71 ) were imaged together and so were iodine ( Z = 53 ), Gd, and holmium ( Z = 67 ). Each phantom was imaged with 1-mm aluminum-filtered 120-kVp cone beam x rays to produce six energy-binned computed tomography (CT) images. Results: K -edge images were reconstructed using a weighted sum of six CT images, which distinguished each contrast agent with a root-mean-square error (RMSE) of 0.29 % and 0.51% for the 0.5% and 5% concentrations, respectively. Minimal cross-contamination in each K -edge image was seen, with RMSE values 0.27 % in vials with no contrast. Conclusion: This is the first preliminary demonstration of simultaneously imaging three similar Z contrast agents with a difference in Z as low as 3.