Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

2009 
Abstract Background: Nocturnal hypoglycemia is a significant problem. From 50% to 75% of hypoglycemia seizures occur at night. Despite the development of real-time glucose sensors (real-time continuous glucose monitor [CGM]) with hypoglycemic alarms, many patients sleep through these alarms. The goal of this pilot study was to assess the feasibility using a real-time CGM to discontinue insulin pump therapy when hypoglycemia was predicted. Methods: Twenty-two subjects with type 1 diabetes had two daytime admissions to a clinical research center. On the first admission their basal insulin was increased until their blood glucose level was <60 mg/dL. On the second admission hypoglycemic prediction algorithms were tested to determine if hypoglycemia was prevented by a 90-min pump shutoff and to determine if the pump shutoff resulted in rebound hyperglycemia. Results: Using a statistical prediction algorithm with an 80 mg/dL threshold and a 30-min projection horizon, hypoglycemia was prevented 60% of the time. ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    22
    References
    117
    Citations
    NaN
    KQI
    []