Rapid estimation of split renal function in kidney donors using software developed for computed tomographic renal volumetry

2011 
Abstract Purpose To evaluate the speed and precision of split renal volume (SRV) measurement, which is the ratio of unilateral renal volume to bilateral renal volume, using a newly developed software for computed tomographic (CT) volumetry and to investigate the usefulness of SRV for the estimation of split renal function (SRF) in kidney donors. Method Both dynamic CT and renal scintigraphy in 28 adult potential living renal donors were the subjects of this study. We calculated SRV using the newly developed volumetric software built into a PACS viewer (n-SRV), and compared it with SRV calculated using a conventional workstation, ZIOSOFT (z-SRV). The correlation with split renal function (SRF) using 99m Tc-DMSA scintigraphy was also investigated. Results The time required for volumetry of bilateral kidneys with the newly developed software (16.7 ± 3.9 s) was significantly shorter than that of the workstation (102.6 ± 38.9 s, p Conclusion The newly developed software for CT volumetry was more rapid than the conventional workstation volumetry and just as accurate, and was suggested to be useful for the estimation of SRF and thus the dominant side in kidney donors.