Do you have to remove a corroded femoral stem

2014 
Abstract Corrosion at the head–neck taper has been recently identified as a cause of adverse local tissue reaction. There are no guidelines concerning removal of fixed femoral components when corrosion is present. The objective of this study is to report the survivorship when a new metal ball is placed on a corroded stem. We examined 86 retrieved femoral heads from metal-on-polyethylene THAs that underwent head and liner exchanges after a minimum 10 years in-vivo and evaluated the subsequent survivorship. There were 7 rerevisions (8.1%) but none were for corrosion-related diagnoses and there was no difference in the survivorship between the 32 THAs with high-grade head taper corrosion and the 54 THAs with low-grade corrosion. We do not currently recommend removing well-fixed femoral stems with corrosion.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    8
    References
    30
    Citations
    NaN
    KQI
    []