Gold Nanoparticle Determination by Inductively Coupled Plasma-Mass Spectrometry, Anodic Stripping Voltammetry, and Flame Atomic Absorption Spectrophotometry

2012 
Gold Nanoparticles (AuNP) were measured by Inductively Coupled Plasma–Mass Spectrometry (ICP-MS), Anodic Stripping Voltammetry (SV), and flame Atomic Absorption Spectrophotometry (AAS). Experiments investigated the relationships between counts per second (ICP-MS), absorbance (AAS), or μA (SV) and Au concentrations in solutions bearing AuNP with sizes of 5, 15, and 50 nm. Similarly the impact of the solution matrix was assessed using deionized water, 1.0 M HNO3, 1.0 M HCl (ICP-MS and AAS), and water containing the bacterium E. coli (∼106 organisms/mL) by all three types of instrumentation. Each instrument yielded linear calibration curves with a signal proportional to Au concentration over the concentration range 0.02 ppm to 1 ppm.The methods were all reliable when biomacromolecules or when organisms such as E.coli existed in the matrix. Comparing the limits of detection for the three methods, ICP-MS and SV were lower than AAS. Comparing cost, SV and AAS were less expensive than ICP-MS. Comparing time requ...