Outcomes of Dissection Angles as Predictor of Restenosis after Drug-Coated Balloon Treatment.

2020 
AIM The predictors of restenosis after endovascular therapy (EVT) with paclitaxel drug-coated balloons (DCBs) have not been clearly established. The present study aimed to investigate the association of post-procedural dissection, as evaluated using intravascular ultrasound (IVUS), with the risk of restenosis following femoropopliteal EVT with paclitaxel DCBs. METHODS In the present single-center retrospective study, 60 de novo femoropopliteal lesions (44 patients) that underwent EVT with DCBs, without bail-out stenting, were enrolled. The primary outcome was 1-year primary patency. Risk factors for restenosis were evaluated using a Cox proportional hazards regression model and random survival forest analysis. RESULTS The 1-year primary patency rate was 57.2% [95% confidence interval, 45%-72%]. IVUS-evaluated post-procedural dissection was significantly associated with the risk of restenosis (P=0.002), with the best cutoff point of 64o [range, 39o-83o]. The random survival forest analysis showed that the variable importance measure of IVUS-evaluated dissection was significantly lower than that of the reference vessel diameter (P<0.001), not different from that of the lesion length (P=0.41), and significantly higher than that of any other clinical feature (all P<0.05). CONCLUSION IVUS-evaluated post-procedural dissection was associated with 1-year restenosis following femoropopliteal EVT with DCB.