Determining the Precision of Dual Energy X-Ray Absorptiometry and Bioelectric Impedance Spectroscopy in the Assessment of Breast Cancer-Related Lymphedema

2013 
Abstract Background: The composition of breast cancer-related lymphedema (BCRL) has been shown to evolve from the initial accumulation of fluid to the development of fibrotic lesions and abnormal fat deposition. Therefore, precise and reliable assessments of BCRL are required to develop accurate staging and management. Although dual energy x-ray absorptiometry (DXA) and bioelectric impedance spectroscopy (BIS) have been used to assess BCRL, no study has evaluated the precision of these two modalities in the same cohort. Methods and Results: We determined the precision of DXA and BIS in lymphedematous (LE) and nonaffected (NA) arms of 24 women with Stage II unilateral BCRL. Precision was calculated from the results of paired bilateral arm measurements obtained from DXA scans measuring fat, lean, and bone mineral masses, BIS measuring extracellular fluid (ECF) and total fluid volume, and circumferential tape measurements (CM) of the arms to calculate the anatomic volume. Precision error was expressed as the...