CT pulmonary angiography using different noise index values with an iterative reconstruction algorithm and dual energy CT imaging using different body mass indices: Image quality and radiation dose.
2017
OBJECTIVE: To investigate the differences in imaging quality and radiation dose in CT pulmonary angiography (CTPA) by using fast-kV switching dual energy CT imaging and 3D Smart mA modulation at different body mass indices (BMIs) and at different noise index (NI) values with an adaptive statistical iterative reconstruction (ASIR) algorithm. METHODS: Four hundred patients who underwent CTPA were equally divided into two groups: A (18.5 kg/m2 ≦ BMI 0.05). The ED in subgroup A4 was 67.12% and 31.53% lower than that in A1 and A2, respectively. In group B, the variables showed no significant differences between the subgroups B3, B1, and B2 (P > 0.05). The ED in subgroup B3 was 50.12% and 35.95% lower than that in B1 and B2, respectively. CONCLUSIONS: Setting different NI values according to BMIs and applying the ASIR algorithm can more effectively reduce the radiation dose in CTPA than in fast-kV switching dual energy CT, while maintaining image quality. Imaging may be performed at NI = 46 in patients with lower BMI (group A) and at NI = 36 in patients with higher BMI (group B).
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
32
References
4
Citations
NaN
KQI