Correlation between glucose concentrations in serum, plasma, and whole blood measured by a point-of-care glucometer and serum glucose concentration measured by an automated biochemical analyzer for canine and feline blood samples

2015 
Objective—To investigate the correlation between glucose concentrations in serum, plasma, and whole blood measured by a point-of-care glucometer (POCG) and serum glucose concentration measured by a biochemical analyzer. Design—Prospective clinical study. Samples—96 blood samples from 80 dogs and 90 blood samples from 65 cats. Procedures—Serum, plasma, and whole blood were obtained from each blood sample. The glucose concentrations in serum, plasma, and whole blood measured by a POCG were compared with the serum glucose concentration measured by a biochemical analyzer by use of the Lin concordance correlation coefficient (ρc) and Bland-Altman plots. Results—For both canine and feline samples, glucose concentrations in serum and plasma measured by the POCG were more strongly correlated with the serum glucose concentration measured by the biochemical analyzer (ρc, 0.98 for both canine serum and plasma; ρc, 0.99 for both feline serum and plasma) than was that in whole blood (ρc, 0.62 for canine samples; ρc, 0...