Flow injection analysis with multiple on-line spectroscopic analysis (UV, IR, 1H-NMR and MS).

2002 
Abstract Studies on the capabilities of flow injection analysis (FIA) combined with on-line characterisation of model compounds via a combination of diode array UV, 1 H-NMR, FT-IR spectroscopy and mass spectrometry are described. Using this combination of spectrometers enabled the on-flow collection of UV, 1 H-NMR, IR and MS for a range of model compounds. Samples were introduced into the system as solutions in deuterium oxide in concentrations ranging from 1.4 to 8.4 mg ml −1 . A sample volume of 100 μl was used for FIA at a flow rate of 1 ml min −1 . From these studies a practical working quantity of ca. 140 μg/sample of analyte was determined which provided characteristic spectra.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    15
    References
    21
    Citations
    NaN
    KQI
    []