Simulation and response of the MART-LIME x-ray detector

1996 
The MART-LIME is a large area x-ray experiment planned to be launched on board the Russian satellite Spectrum X-Gamma, as the high energy imager of a complement of broad band co- aligned x-ray telescopes. The energy range covered is 5 - 150 keV with an angular resolution of 8.6 arcminutes. The final detector configuration is now in its testing phase and includes the high pressure window comprising the 6 by 6 degree collimator, and the multiwire proportional counter (MWPC). The response to x-ray sources was investigated during the tests carried out at the Daresbury Laboratory (Warrington, UK) facilities The MWPC was filled up by a xenon-argon-isobutane gas mixture in order to evaluate the efficiency of the detector and in particular its linearity over the whole approximately 2,000 cm 2 sensitive area. At the same time the various parts of the apparatus have been simulated by using a Monte Carlo program. Results on the detector response and simulations are presented.