Transcervical carotid stenting with dynamic flow reversal demonstrates embolization rates comparable to carotid endarterectomy

2016 
Purpose: To evaluate a series of patients treated electively with carotid endarterectomy (CEA), transfemoral carotid artery stenting with distal filter protection (CASdp), and transcervical carotid stenting with dynamic flow reversal (CASfr) monitored continuously with transcranial Doppler (TCD) during the procedure to detect intraoperative embolization rates. Methods: Thirty-four patients (mean age 67.6 years; 24 men) with significant carotid stenosis underwent successful TCD monitoring during the revascularization procedure (10 CEA, 8 CASdp, and 16 CASfr). Ipsilateral microembolic signals were segregated into 3 phases: preprotection (until internal carotid artery cross-shunted or clamped if no shunt was used, filter deployed, or flow reversal established), protection (until clamp/shunt was removed, filter retrieved, or antegrade flow reestablished), and postprotection (after clamp/shunt or filter removal or restoration of normal flow). Results: CASdp showed higher embolization rates than CEA or CASfr in...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    36
    References
    37
    Citations
    NaN
    KQI
    []