The influence of sample porosity on refractive index in THz non-destructive testing

2021 
Terahertz time-domain spectroscopy has been widely applied in performing dielectric analysis for various materials. However, air voids trapped in samples will significantly affect the characterization. In this study, the refractive index of mixture samples consisting of iron trioxide and polytetrafluoroethylene in five different mass ratios were measured with terahertz time-domain spectroscopy. In order to extract the intrinsic refractive index of iron trioxide, the effective medium model of CRI was then applied to remove refractive index of PTFE in the sample. The extracted refractive indices were presented as a variable parameter decrease with the increase of iron trioxide content, which also corresponds to the increase of air voids in the tablets. The correlation between trapped air voids and analyte composition roots from the coarse rust particles used in pellets compression. This study shows that the influence of air porosity should be considered when terahertz waves are utilized to characterize dielectric property of coarse material.