Computer-Assisted Three-Dimensional Morphology Evaluation of Intracranial Aneurysms

2018 
Objective Precise morphologic evaluation is important for intracranial aneurysm (IA) management. At present, clinicians manually measure the IA size and neck diameter on 2-dimensional (2D) digital subtraction angiographic (DSA) images and categorize the IA shape as regular or irregular on 3-dimensional (3D)-DSA images, which could result in inconsistency and bias. We investigated whether a computer-assisted 3D analytical approach could improve IA morphology assessment. Methods Five neurointerventionists evaluated the size, neck diameter, and shape of 39 IAs using current and computer-assisted 3D approaches. In the computer-assisted 3D approach, the size, neck diameter, and undulation index (UI, a shape irregularity metric) were extracted using semiautomated reconstruction of aneurysm geometry using 3D-DSA, followed by IA neck identification and computerized geometry assessment. Results The size and neck diameter measured using the manual 2D approach were smaller than computer-assisted 3D measurements by 2.01 mm ( P P Conclusions Computer-assisted 3D morphology analysis can improve accuracy and consistency in measurements compared with manual 2D measurements. It can also more reliably quantify shape irregularity using the UI. Future application of computer-assisted analysis tools could help clinicians standardize morphology evaluations, leading to more consistent IA evaluations.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    22
    References
    11
    Citations
    NaN