Dose reduction in orthodontic lateral cephalography: dosimetric evaluation of a novel cephalographic thyroid protector (CTP) and anatomical cranial collimation (ACC)

2015 
Objectives:To test the dose-reducing capabilities of a novel thyroid protection device and a recently introduced cranial collimator to be used in orthodontic lateral cephalography.Methods:Cephalographic thyroid protector (CTP) was designed to shield the thyroid while leaving the cervical vertebrae depicted. Using a RANDO® head phantom (The Phantom Laboratory, Salem, NY) equipped with dosemeters and a Proline XC (Planmeca, Helsinki, Finland) cephalograph, lateral cephalograms were taken, and the effective dose (ED) was calculated for four protocols: (1) without shielding; (2) with CTP; (3) with CTP and anatomical cranial collimator (ACC); and (4) with a thyroid collar (TC).Results:The ED for the respective protocols was (1) 8.51; (2) 5.39; (3) 3.50; and (4) 4.97 µSv. The organ dose for the thyroid was reduced from 30.17 to 4.50 µSv in Protocols 2 and 3 and to 3.33 µSv in Protocol 4.Conclusions:The use of just the CTP (Protocol 2) resulted in a 36.8% reduction of the ED of a lateral cephalogram. This was co...