Transmyocardial Laser Revascularization: Epicardial ECG Detection Provides Efficient R-Wave Triggering during Mobilization of the Heart

2003 
Objective: In order to achieve an accurate intraoperative ECG detection, a new technique in detecting the trigger-signal was developed. In contrast to the traditional three-lead ECG-configuration, the left leg electrode was connected to a transient epicardial pacemaker electrode on the left-ventricular surface. Background Data: The Holmium:YAG-Laser for Transmyocardial Laser Revascularization (TMLR) is R-wave-triggered, providing the release of energy only during the refractory period of the heart cycle. However, an exact ECG-triggering during mobilization of the apex and/or posterior wall is difficult to achieve by using conventional ECG-configuration, therefore increasing the risk for mistriggering and induction of arrhythmias during TMLR. Materials and Methods: Two groups of patients, all undergoing stand alone TMLR-procedures via left minithoracotomy, were compared. Ten patients were operated with the conventional ECG configuration (group 1) and ten patients with the modified epicardial ECG configurat...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    1
    Citations
    NaN
    KQI
    []