Impact of managing affected results in haemolysed samples of an infant-maternity hospital using an unconventional approach.

2021 
BACKGROUND The management of affected results in haemolysed samples (HS) is debated. In an infant-maternity setting, for reporting interfered test results, we provided the result itself, the degree of haemolysis (as free haemoglobin concentration), and a warning recommending sample recollection. We investigated the impact of this approach on sample quality and clinicians' decision-making. METHODS Free haemoglobin was measured on Beckman Coulter AU680 as haemolytic index. We estimated the total HS number, the clinical wards more affected by HS, the most interfered analytes, and the retesting rate of interfered tests, by comparing data from Apr-Dec 2017, the period just after the introduction of the new policy, vs. Apr-Dec 2018. RESULTS One year after the new report introduction, a significant HS decrease (5.8% vs. 7.8%, P < 0.001) was detected, together with a reduction of the frequency by which haemolysis affected results. The most affected wards, i.e., Paediatric and Neonatal Intensive Care Units, showed an improvement in sample quality (HS rate, 30.6% to 16.1%, P < 0.001, and 25.2% to 20.9%, P = 0.048, respectively). We noted a significant decrease in retesting after an alerted result for aspartate aminotransferase, magnesium, potassium, conjugated bilirubin, and lactate dehydrogenase. CONCLUSIONS Our approach led to a HS decrease, suggesting that the provided report could be a driving force for improvement of phlebotomy quality, also helping clinicians in deciding if retesting is essential or not.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    0
    Citations
    NaN
    KQI
    []