Diagnostic efficiency of troponin T measurements in acute myocardial infarction.

1991 
BACKGROUNDThe present study was designed to evaluate the efficiency of a newly developed troponin T enzyme immunoassay for the detection of acute myocardial infarction.METHODS AND RESULTSThe study comprised 388 patients admitted with chest pain and suspected myocardial infarction and 101 patients with skeletal muscle damage and additional suspected myocardial cell damage. Troponin T was elevated to more than twice the analytical sensitivity of the assay (0.5 microgram/l) in all patients with non-Q wave (range, 1.2-5 micrograms/l) and Q wave infarction (range, 3-220 micrograms/l). Troponin T appeared in serum as early as 3 hours after onset of pain in 50% of the patients and remained elevated in all patients for more than 130 hours, revealing release kinetics of both free cytosolic and structurally bound molecules. The diagnostic efficiency of troponin T was superior to that of creatine kinase-MB (98% versus 97%) and remained at 98% until 5.5 days after admission, if patients with unstable angina were excl...