Influence of body position, food and beverage consumption on BIS measurements

2010 
Continuous monitoring of fluid changes using bioimpedance spectroscopy (BIS) during hemodialysis could help to predict hypotensive complications and extend the patient's life. Food and beverage consumption during the treatment may influence the measurements and the calculated fluid removal. In the present article the change observed in whole body and segmental (knee-to-knee, abdomen) BIS measurements following a sequence similar to the one of dialysis treatment (lying down, sitting and eating, lying down) on healthy subjects is presented. The measurements have been performed using a commercial bioimpedance device with a frequency range of 5 kHz to 1 MHz. Knee-to-knee measurements seem to be less sensitive to these influences, compared to the standard whole body and the alternative abdomen BIS measurements. The results indicate that the individual influence of both body posture and food and beverage consumption may be superposed when combined.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    9
    References
    4
    Citations
    NaN
    KQI
    []