DESIGN AND CLINICAL RESULTS OF SUTURE-BUTTON ANKLE SYNDESMOSIS FIXATION

2008 
A new apparatus and technique of syndesmosis fixation is tested in a prospective clinical study. Buttons on both sides of the ankle anchor a strong suture under tension following syndesmosis reduction. This syndesmosis suture acts like a tightrope when under tension. Implantation is simple with a minimally invasive technique, as the medial side is not opened. It allows physiological micromotion whilst resisting diastasis, does not require routine removal, and allows patients to weight-bear earlier. Sixteen patients with Weber C ankle fractures with a syndesmosis diastasis underwent suture-button fixation and the results compared to 16 consecutive patients with syndesmosis screw fixation. Patients were, in effect, quasi-randomised according to surgeon preference. Mean A,O,F,A,S, ankle scores were significantly better in the suture-button group at three months post-op (91 vs 80, p=0.01, unpaired t-test) and at twelve months (93 vs 83, p=0.04, unpaired t-test). Return to work was also significantly faster (2.6 months vs 4.6 months, p=0.02, unpaired t-test). No suture-buttons required implant removal. Axial CT scanning at three months showed implants to be intact with maintenance of reduction, as compared to the uninjured contralateral side. Suture-button syndesmosis fixation is simple, safe and effective. It has shown improved outcomes and faster rehabilitation, without needing routine removal. Although the apparatus design may undergo further refinement, we believe this technique will become the treatment of choice in Weber C ankle fractures with a syndesmosis injury.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []