A simple dose gradient measurement tool to complement the conformity index

2006 
✓A dose gradient index (GI) is proposed that can be used to compare treatment plans of equal conformity. The steep dose gradient outside the radiosurgical target is one of the factors that makes radiosurgery possible. It therefore makes sense to measure this variable and to use it to compare rival plans, explore optimal prescription isodoses, or compare treatment modalities. The GI is defined as the ratio of the volume of half the prescription isodose to the volume of the prescription isodose. For a plan normalized to the 50% isodose line, it is the ratio of the 25% isodose volume to that of the 50% isodose volume. The GI will differentiate between plans of similar conformity, but with different dose gradients, for example, where isocenters have been inappropriately centered on the edge of the target volume. In a retrospective series of 50 dose plans for the treatment of vestibular schwannoma, the optimal prescription isodose was assessed. A mean value of 40% (median 38%, range 30–61%) was calculated, not...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    346
    Citations
    NaN
    KQI
    []