Design and analysis of a burst strength device for testing vascular grafts

2019 
The design and analysis of a device to measure the burst strength (strength under a state of pure radial internal pressure) and compliance of vascular grafts and flexible pressurized tubes is presented. The device comprises three main sections, viz., a clean air-dry pressure controller, a test specimen holder, and automated software for control and data collection. Air pressure is controlled by means of a valve and a dedicated mechanism allowing reaching up to 120 psi in increments of 1 psi, and recording pressure changes with 0.04 psi resolution. The circumferential strain is determined by measuring the radial displacement of the vascular graft using an optical arrangement capable of determining a maximum radial displacement of 10 mm with 0.02 mm resolution. The instrument provides a low uncertainty in compliance (±0.32%/100 mm Hg−1) and burst strength measurements. Due to its simplicity, the device can easily be reproduced in other laboratories contributing to a dedicated instrument with high resolution...