Comparison of pulsed Doppler and thermodilution methods for measuring cardiac output in critically ill patients.
1987
: We obtained 145 consecutive cardiac output measurements in 38 critically ill patients, using the invasive thermodilution and the noninvasive pulsed Doppler methods. The mean thermodilution cardiac output (TDCO) was 5.7 +/- 1.87 L/min and the mean pulsed Doppler cardiac output (PDCO) was 5.16 +/- 1.66 L/min. The mean difference between the two measurements was 0.51 L/min with an SD greater than 1.6 L/min, reflecting the scattering of results. The overall correlation coefficient was .58. The intercepts were large and the regression equation some way from the line of equal values (TDCO = 2.28 + 0.66 PDCO). When the results were analyzed according to diagnosis or by group experience, there were some differences in the bias of the estimate; however, the SD of the difference between methods was greater than one liter/min in all groups. Thus, the pulsed Doppler method failed to estimate accurately TDCO in critically ill patients.