The vascular cooling effect in hepatic multipolar radiofrequency ablation leads to incomplete ablation ex vivo

2016 
AbstractPurpose: Major limitations of conventional RFA are vascular cooling effects. However, vascular cooling effects are supposed to be less pronounced in multipolar RFA. The objective of this ex vivo study was a systematic evaluation of the vascular cooling effects in multipolar RFA.Materials and methods: Multipolar RFA with three bipolar RFA applicators was performed ex vivo in porcine liver (applicator distance 20 mm, energy input 40 kJ). A saline-perfused glass tube (‘vessel’) was placed parallel to the applicators in order to simulate a natural liver vessel. Five applicator-to-vessel geometries were tested. A liquid-filled glass tube without perfusion was used as a dry run. Ablations were orthogonally cut to the applicators at a defined height. Cooling effects were analysed qualitatively and quantitatively along these cross sectional areas.Results: Thirty-six ablations were performed. A cooling effect could be seen in all ablations with perfused vessels compared to the dry run. While this cooling e...