A novel sax-stent method in treatment of ascending aorta and aortic arch aneurysms evaluated by finite element simulations

2017 
Summary Objectives A novel stent method to simplify treatment of proximal ascending aorta and aortic arch aneurysms was developed and investigated by finite element analysis. Therapy of ascending aortic and aortic arch aneurysms is difficult and challenging and is associated with various complications. Methods A 55 mm wide × 120 mm long stent was designed without the stent graft and the stent was deployed by an endovascular method in a virtual patient-specific aneurysm model. The stress-strain analysis and deployment characteristics were performed in a finite element analysis using the Abaqus software. Results The stent, when embedded in the aortic wall, significantly reduced aortic wall stresses, while preserving the side coronary ostia and side branches in the aortic arch. When tissue growth was modeled computationally over the stent struts the wall stresses in aorta was reduced. This effect became more pronounced when increasing the thickness of the tissue growth. There were no abnormal stresses in the aorta, coronary ostium and at the origin of aortic branches. The stent reduced aneurysm expansion cause by hypertensive condition from 2 mm without stenting to 1.3 mm after stenting and embedding. Conclusion In summary, we uncovered a simple treatment method using a bare nitinol stent without stent graft in the treatment of the proximal aorta and aortic arch aneurysms, which could eventually replace the complex treatment methods for this disease.
    • Correction