Comparison of the SITA Faster-a new visual field strategy with SITA Fast strategy.

2021 
AIM To compare visual field defects using the Swedish Interactive Thresholding Algorithm (SITA) Fast strategy with SITA Faster strategy, a newly developed time-saving threshold visual field strategy. METHODS Ninety-three participants (60 glaucoma patients and 33 normal controls) were enrolled. One eye from each participant was selected randomly for the study. SITA Fast and SITA Faster were performed using the 24-2 default mode for each test. The differences of visual field defects between the two strategies were compared using the test duration, false-positive response errors, mean deviation (MD), visual field index (VFI) and the numbers of depressed test points at the significant levels of P 0.05). Correlation analysis showed a high correlation for MD (r=0.986, P<0.001) and VFI (r=0.986, P<0.001) between the two strategies. Bland-Altman analysis showed great agreement between the two strategies. CONCLUSION SITA Faster, which saves considerable test time, has a great test quality comparing to SITA Fast, but may be not directly interchangeable.