The Optimal Cell Salvage Settings to Maximize Hematocrit and Minimize Potassium Using the Cobe BRAT2 Autologous Blood Recovery Unit.

2021 
Objective The objective was to determine the optimal cell saver device settings (infusion rate and wash rate) at which hematocrit is preserved and potassium and lactate are removed from banked red blood cells (RBC). Design Red cells were washed using the Cobe BRAT 2 Autologous Blood Recovery Unit and sampled for electrolyte composition and hematocrit pre- and postwash. Setting This was a single-center study. Interventions Red cells were washed using six infusion rates (100-1,000 mL/min) and six wash rates (100-1,000 mL/min) for a total of 36 combinations. Hematocrit, potassium, glucose, and lactate were evaluated before and after washing. Measurements and Main Results At wash rates ≤400 mL/min, hematocrit increased independent of infusion rate. At wash rates ≥400 mL/min, slower infusion rates were associated with higher hematocrit compared to faster infusion rates (p Conclusion Red-cell washing produces higher hematocrit and lower potassium as infusion rate and wash rate decrease. A 340-mL unit of RBC can be processed in 4.26 minutes without loss of hematocrit or an increase in potassium when both infusion and wash rates are set to 400 mL/min.