Multichannel magnetic stimulation system design considering mutual couplings among the stimulation coils

2004 
We introduce some simulation and experiment results of the multichannel magnetic stimulator development that has been carried out as an initial attempt to realize a multichannel functional magnetic stimulator. For efficient functional magnetic stimulations, precise spatial localization of stimulation sites without any movements of the stimulation coils is very important. We have found that the mutual coupling effect among the adjacent stimulation coils in the coil array has to be considered in the determination of the charge voltages in some coil array configurations. Experimental results obtained with a 4-channel magnetic stimulator are presented.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    34
    Citations
    NaN
    KQI
    []