A Comparison of Proximal Tibia, Distal Femur, and Proximal Humerus Infusion Rates Using the EZ-IO Intraosseous Device on the Adult Swine (Sus scrofa) Model

2013 
AbstractObjective. We sought to compare the flow rates of the proximal tibia, the distal femur, and the proximal humerus using high pressure (greater than 300 mmHg) through an intraosseous (IO) infusion needle in an adult swine model. Methods. We performed a prospective interventional study in 11 swine. After placement of central vein and arterial lines, blood was removed via the central line until the animal's mean arterial pressure decreased 25% from the recorded baseline. We inserted a 25-mm IO needle into the proximal tibia and 45-mm needles into the distal femur and proximal humerus. All extremities were utilized in each study animal. We infused normal saline at each site for 10 minutes with a pressure bag inflated to the highest achievable pressure (greater than 300 mmHg) as measured at the infusion site with a calibrated portable inline pressure gauge. We measured the volume of normal saline remaining and we calculated infusion rates for each site. We then compared infusion flow rates between the t...