Non-invasive measurement of reservoir pressure parameters from brachial-cuff blood pressure waveforms.

2018 
Reservoir pressure parameters [eg, reservoir pressure (RP) and excess pressure (XSP)] are biomarkers derived from blood pressure (BP) waveforms that have been shown to predict cardiovascular events independent of conventional cardiovascular risk markers. However, whether RP and XSP can be derived non-invasively from operator-independent cuff device measured brachial or central BP waveforms has never been examined. This study sought to achieve this by comparison of cuff reservoir pressure parameters with intra-aortic reservoir pressure parameters. 162 participants (aged 61 ± 10 years, 72% male) undergoing coronary angiography had the simultaneous measurement of cuff BP waveforms (via SphygmoCor XCEL, AtCor Medical) and intra-aortic BP waveforms (via fluid-filled catheter). RP and XSP derived from cuff acquired brachial and central BP waveforms were compared with intra-aortic measures. Concordance between brachial-cuff and intra-aortic measurement was moderate-to-good for RP peak (36 ± 11 vs 48 ± 14 mm Hg, P vs 24 ± 9 mm Hg, P vs 46 ± 14 mm Hg, P vs 24 ± 9 mm Hg, P  < 0.001; ICC 0.12, 95% CI: -0.13 to 0.31). In conclusion, both brachial-cuff and central-cuff methods can reasonably estimate intra-aortic RP, whereas XSP can only be acceptably derived from brachial-cuff BP waveforms. This should enable widespread application to determine the clinical significance, but there is significant room for refinement of the method.