Determination of total homocysteine in dried blood spots using high performance liquid chromatography for homocystinuria newborn screening
2004
Background: The most widely used method for newborn screening for homocystinuria (HCU) is a semiquantitative bacterial inhibition assay for measuring methionine concentration in dried blood spots (DBS). Because this method has resulted in a number of missed cases due to many factors, we developed a high performance liquid chromatography (HPLC) method with fluorescence detection to measure total homocysteine (tHcy) in DBS which might be useful for newborn screening for HCU.
Methods: One disk of DBS 3 mm in diameter was sonicated in 10 min. The extract was reduced with dithioerythritol and was derivatized with 4-aminosulfonyl-7fluoro-2,1,3-benzoxadiazole before injection into HPLC.
Results: This method showed good linearity (r = 0.996), precision (coefficient of variation range 2.7−5%), and excellent correlation coefficient between DBS and serum tHcy, both in control (r = 0.932) and patient samples (r = 0.952). By this method, the mean tHcy concentration in DBS of preterm newborns, full-term newborns, and adults was 1.4 ± 1.0, 2.5 ± 1.6, and 4.9 ± 1.5 µmol/L, respectively. The mean tHcy DBS concentration in two cases of cystathionine-β-synthase deficiency and one case of 5,10-methylentetrahydrofolate reductase deficiency was 22.7 ± 2.88, 29.3 ± 1.90, and 41.3 µmol/L, respectively.
Conclusions: The present method, which is rapid, user friendly and reliable, seems applicable to newborn screening of HCU in place of methionine measurement.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
14
References
20
Citations
NaN
KQI