Thermotherapy: feasibility study using a single focussed ultrasound transducer

1999 
Feasibility studies on tissue lesioning using high intensity focussed ultrasound (HIFU) thermotherapy were carried out. A single strongly focussed transducer was used. The transducer has a diameter of 50mm, a centre resonance frequency of 1MHz and a focal distance of 50mm. Experiments were carried out both in vitro on pig muscle samples and in vivo on a pig (exposures on the thigh and on the liver). Different types of burst pulse signals were used, the best results were obtained with 1-2s burst pulses, with a repetition period of 10s. The risk for skin damage increases for longer burst pulses. Lesions at the focus were obtained both in the experiments in vitro as well as in vivo . The size and the shape of the lesions in vivo resemble those obtained in vitro . The results of the experiments show that tissue lesions can be obtained at deep locations (4.5cm in the tissue) with the transducer. The lesions were well discriminated from the surrounding tissue and were ellipsoid- or drop-shaped. The lesion size ...