Long-time measurements of line-integrated plasma electron density using a two-color homodyne optical fiber interferometer.

2021 
A two-color homodyne Mach–Zehnder (M–Z) optical fiber interferometer with wavelengths of 1.55 and 1.31 µm was developed for long-time measurement of line-integrated plasma electron density. A novel phase difference demodulation algorithm based on a single 3 × 3 optical coupler was implemented in a two-color optical fiber interferometer scheme for the first time. Our laboratory tests showed that this new optical fiber interferometer could determine the phase shift due to the low-frequency ambient vibration and could maintain high phase resolution measurement. The resolution of the new interferometer was less than 0.04 rad in 1000 s, corresponding to a line-averaged electron density of less than 1.0 × 1019 m−2. Actual plasma discharge experiments performed on KTX-CTI, which is a new compact torus injector (CTI) constructed at the Keda Torus eXperiment (KTX), showed that this interferometer has excellent several-second stability.