Comparison of two different methods for measurement of phenylalanine in dried blood spots.
2000
Phenylketonuria is an inherited metabolic disorder caused by a defect in the hydroxylation of phenylalanine. Newborn screening is crucial for the diagnosis and treatment of this disease. A phenylalanine dehydrogenase-coupled enzymatic assay (Quantase) in microtiter plates for the screening of phenylketonuria was evaluated and compared with our routine method based on the modified fluorometric McCaman method. The test exhibited a linear calibration curve with a good slope as well as sufficient imprecision (< 10%), recovery (99.23 ± 4.86%) and limit of detection (54.5 μmol/l). One hundred and ninety dried blood spots were analysed by this enzymatic method and compared with McCaman's. Although Quantase (Teknovas, Bilbao, Spain) showed a phenylalanine mean level in dried blood spot 18.2 μmol/l higher than that obtained with our routine method, the agreement between both techniques was considered acceptable.