Oxygen nipple and nut (Christmas tree) adaptor contamination rates and decontamination with disinfecting wipes
2020
OBJECTIVE: Different manufacturers recommend different levels of disinfection for oxygen nipple and nut adaptors, also known as Christmas-tree adaptors (CTAs). We aimed to determine the bacterial contamination rates of CTAs before and after clinical use and whether disinfection wipes effectively eliminate bacteria from CTAs. METHODS: CTAs were swabbed for bacteria directly from the shipment box or after use in a medical intensive care unit to determine levels of contamination. CTAs were also inoculated in the laboratory with a variety of bacteria and disinfected with either 0.5% hydrogen peroxide (Oxivir 1) or 0.25% tetra-ammonium chloride with 44.50% isopropyl alcohol (Super Sani-Cloth), and the effectiveness of each wipe was determined by comparing the bacterial recovery before and after disinfection. RESULTS: CTAs exhibit low levels of bacterial burden before and after clinical use. Both disinfecting wipes were effective at removing bacteria from the CTAs. CONCLUSIONS: Low-level disinfection of CTAs is appropriate prior to redeployment in the clinical setting.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
4
References
0
Citations
NaN
KQI