Design and verification of an external radiobiological beam port on a 16.5 MeV GE PETtrace proton cyclotron

2019 
PURPOSE: Protons and heavy ions are considered to be ideal particles for use in external beam radiotherapy due to the superior properties of the dose distribution. While a photon (x-ray) beam delivers considerable dose to healthy tissues around the tumor, a proton beam that is delivered with sufficient energies has: a low entrance dose (the dose in front of the tumor); a high-dose region within the tumor, known as the Bragg peak; and, no exit dose beyond the tumor. Proton therapy is the next major step in advancing radiotherapy treatment. The purpose of this project was to adapt an existing radioisotope production cyclotron, a General Electric (GE) PETtrace, to enable radiobiological studies using proton beams. During routine use the PETtrace delivers 16.5 MeV protons to target with beam currents in the range of 10-100 microA resulting in dose rates in the order of kGy/s. To achieve the aim of the project the dose rate had to be reduced to the Gy/min range, without attenuating the proton energy below 5 MeV. This paper covers the design, construction and validation of the beam port. METHODS: Monte Carlo simulations were performed, using GEANT4, SRIM and PACE4 to design the beam port and optimize its components. Once the beam port was fabricated, validation experiments were performed using EBT3 and HD-V2 Gafchromic films, and a Keithley 6485 picoampere meter. RESULTS AND CONCLUSION: The external beam port was successfully modeled, designed and fabricated. By using a 0.25 mm thick gold foil and a brass pin-hole collimator the beam was spread from a narrow full beam diameter of 10 mm to a wide beam with a 5% flatness area in the center of the beam that had a diameter of ~20 mm. In using this system the dose rate was reduced from kGy/s to ~30 Gy/min.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    0
    Citations
    NaN
    KQI
    []