Studying the effect of noise on the performance of 2D and 3D texture measures for quantifying the trabecular bone structure as obtained with high resolution MR imaging at 3 Tesla

2008 
3.0 Tesla MRI devices are becoming popular in clinical applications since they render images with a higher signal-tonoise ratio than the former 1.5 Tesla MRI devices. Here, we investigate if higher signal-to-noise ratio can be beneficial for a quantitative image analysis in the context of bone research. We performed a detailed analysis of the effect of noise on the performance of 2D morphometric linear measures and a 3D nonlinear measure with respect to their correlation with biomechanical properties of the bone expressed by the maximum compressive strength. The performance of both 2D and 3D texture measures was relatively insensitive to superimposed artificial noise. This finding suggests that MR sequences for visualizing bone structures at 3T should rather be optimized to spatial resolution (or scanning time) than to signal-to-noise ratio.