Phase-contrast imaging with a compact x-ray light source: system design

2017 
X-ray phase-contrast imaging (XPCI) overcomes the problem of low contrast between different soft tissues achieved in conventional x-ray imaging by introducing x-ray phase as an additional contrast mechanism. This work describes a compact x-ray light source (CXLS) and compares, via simulations, the high quality XPCI results that can be produced from this source to those produced using a microfocus x-ray source. The simulation framework is first validated using an image acquired with a microfocus-source, propagation-based XPCI (PB-XPCI) system. The phase contrast for a water sphere simulating a simple cyst submersed in muscle is evaluated and the evolution of PB-XPCI signal as the object to detector distance is increased is demonstrated. The proposed design of a PB-XPCI system using the CXLS is described and simulated images of a coronary artery compared between CXLS and microfocus source PB-XPCI systems. To generate images with similar noise levels, a microfocus source would require a 3000 times longer exposure than would the CXLS. We conclude that CXLS technology has the potential to provide high-quality XPCI in a medical environment using extremely short exposure times relative to microfocus source approaches.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    32
    References
    1
    Citations
    NaN
    KQI
    []