Comparison of observer reliability of three-dimensional cephalometric landmark identification on subject images from Galileos and i-CAT cone beam CT

2013 
Objectives:Recently, there has been increasing interest in the use of cone beam CT (CBCT) for three-dimensional cephalometric analysis and craniofacial reconstruction in orthodontic and orthognathic surgical treatment planning. However, there is a need to redefine the cephalometric landmarks in three dimensional cephalometric analysis and to demonstrate the reproducibility of landmark identification on the type of CBCT machine being used.Methods:CBCT images of 20 subjects aged 15–25 years were selected, ten each from Galileos® (Sirona Dental Systems Inc., Bensheim, Germany) and Next Generation i-CAT® (Imaging Sciences International, Hatfield, PA). 2 observers located 18 landmarks on each subject twice using Dolphin-3D v. 11 software (Dolphin Imaging and Management Systems, Chatsworth, CA). Inter- and intraobserver reliability was assessed using Euclidean distances and linear mixed models.Results:Overall, the intra- and interobserver reliability was excellent for both machines. The landmarks Gonion, Nasion...