Biosensors Acting in Injection Mode

2021 
This chapter numerically investigates the sensitivity of an amperometric biosensor acting in the flow injection mode when the biosensor contacts an analyte for a short time. The analytical system is modeled by non-stationary reaction–diffusion equations containing a nonlinear term related to the Michaelis–Menten kinetics of an enzymatic reaction. At first, the biosensor action is modeled by a mono-layer mono-enzyme model assuming no external diffusion limitation. Then, the model is extended to a two-compartment model by adding an outer diffusion layer. The biosensor operation is analysed with a special emphasis to the conditions at which the biosensor sensitivity can be increased and the calibration curve can be prolonged by changing the injection duration, the permeability of the external diffusion layer, the thickness of the enzyme layer and the catalytic activity of the enzyme. The apparent Michaelis constant is used as a main characteristic of the sensitivity and the calibration curve of the biosensor. The numerical simulation was carried out using the finite difference technique.