Adherent Cell Assay Results Affected by Variable z-Position Mixing

2010 
We demonstrate that modifying mixing dynamics after addition of organic solute into aqueous buffers dramatically affects cell morphology and protein expression. Variable z-position (VZP) or varying the height of aspiration and dispense positions during mixing eliminates artifactual effects. Here, we tested 4 adherent cell types and show effects of VZP on quantitative imaging, protein expression, viability, and morphology. The result: The quantitation of cytoplasmic fluorescence within the fields of interest of the phalloidin-actin stain assay improved by 47% and fluorescence variability emitted by cells expressing green fluorescence protein (GFP) fusion proteins decreased by 15%. Assays that perform measurement by averaged reading of the entire well are somewhat susceptible. For example, protein production decreased 8% on the hypoxia response element (HRE)-luciferase assay. VZP did not affect quantitative cell viability, deviate the half maximal effective dose concentration (EC50) values or alter expected...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    0
    Citations
    NaN
    KQI
    []