Relationship of Laser-Doppler-Flow and coronary perfusion and a concise update on the importance of coronary microcirculation in donor heart machine perfusion.

2021 
BACKGROUND Machine perfusion (MP) is a novel method for donor heart preservation. The coronary microvascular function is important for the transplantation outcome. However, current research on MP in heart transplantation focuses mainly on contractile function. OBJECTIVE We aim to present the application of Laser-Doppler-Flowmetry to investigate coronary microvascular function during MP. Furthermore, we will discuss the importance of microcirculation monitoring for perfusion-associated studies in HTx research. METHODS Porcine hearts were cardioplegically arrested and harvested (Control group, N = 4). In an ischemia group (N = 5), we induced global ischemia of the animal by the termination of mechanical ventilation before harvesting. All hearts were mounted on an MP system for blood perfusion. After 90 minutes, we evaluated the effect of coronary perfusion pressures from 20 to 100 mmHg while coronary laser-doppler-flow (LDF) was measured. RESULTS Ischemic hearts showed a significantly decreased relative LDF compared to control hearts (1.07±0.06 vs. 1.47±0.15; p = 0.034). In the control group, the coronary flow was significantly lower at 100 mmHg of perfusion pressure than in the ischemia group (895±66 ml vs. 1112±32 ml; p = 0.016). CONCLUSIONS Laser-Doppler-Flowmetry is able to reveal coronary microvascular dysfunction during machine perfusion of hearts and is therefore of substantial interest for perfusion-associated research in heart transplantation.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    0
    Citations
    NaN
    KQI
    []