We have been developing an electrohydraulic total artificial heart (EHTAH) system. The system consists of diaphragm blood pumps, an abdominally placed energy converter, an internal controller, a transcutaneous energy transfer (TET) system, a transcutaneous optical information transfer system, and internal and external lithium-ion (Li-ion) batteries. The energy converter was optimized to obtain better oil transfer. Maximum cardiac output and efficiency of the EHTAH were increased from 8 L/min to 10 L/min and from 10% to 12%, respectively. The volume of the energy converter was reduced from 280 to 210 ml. The pumping unit was successfully implanted in 68-85 kg calves without anatomic problems, and the calves survived up to 10 days with good circulatory results. The maximum temperature rise of the implanted energy converter was only 1°C. Stable performance of the TET system was confirmed in goats for more than 1 month. DC-DC energy transfer efficiency with 20 W of energy transmission remained within the range of 80% to 85%, and no significant temperature rise was observed in the implanted circuit. The internal Li-ion battery was also evaluated in a goat, and the maximum temperature rise during the charging period was 1.5°C, while the charging and discharging times were 72 and 58 min, respectively. We conclude that our system has progressed in its development as a practical implantable system.