Development of an Analytical Method for Simultaneous Determinationof Quinolones and Tetracyclines in Livestock and Fishery Products

2021 
In this study, we developed an analytical method for simultaneous determination of 14 quinolones and 4 tetracyclines in livestock and fishery products using LC-MS/MS. The analytes were extracted from food samples with citrate buffer (containing EDTA)-methanol-acetonitrile (3 : 1 : 1, v/v/v) in the presence of n-hexane, and the extract was purified with an Oasis PRiME HLB cartridge column. It was suggested that this analytical method can also extract analytes from solid samples containing fat by using n-hexane. In addition, using methanol-acetonitrile (3 : 7, v/v) containing 0.1 vol% formic acid as an eluent from the cartridge column, the purification effect could be improved, while minimizing the impairment of the recovery rate. As a result of the validation using six types of food samples, trueness (accuracy) was 70.6%-113.8%, the RSD of repeatability was 9.0% or less, and the RSD of within-laboratory reproducibility was 15.5% or less. Using this approach, the standard values mentioned in the Japanese guideline were successfully met.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []