Long-Term Performance of Right Ventricular Implantable Cardioverter-Defibrillator Leads in Arrhythmogenic Right Ventricular Cardiomyopathy and Hypertrophic Cardiomyopathy

2020 
Patients with arrhythmogenic right ventricular cardiomyopathy (ARVC) and hypertrophic cardiomyopathy (HCM) implanted with implantable cardioverter-defibrillators (ICDs) may show a large decrease in R-wave amplitude during long-term follow-up. However, it is unclear whether this decrease is higher in these patients than in those without structural heart disease. This study investigated ICD-lead intracardiac parameters over a long duration in patients with ARVC and HCM and compared these parameters with those of a control group. We included 50 patients (mean age, 55.2 +/- 17.2 years; 26% female) with ICD leads in the right ventricular apex, and compared 7 ARVC and 14 HCM patients with 29 control patients without structural heart disease. ICD-lead parameters, including R-wave amplitude, pacing threshold, and impedance during follow-up, were compared. The difference in these parameters between the time of implantation and year 5 were also compared. There were no significant differences in R-wave amplitude at implantation among the 3 groups. The change in R-wave amplitude between the time of implantation and year 5 was significantly greater in the ARVC group (-3.3 +/- 5.4 mV, P = 0.012) in comparison to the control group (1.3 +/- 2.8 mV); the HCM group showed no significant difference (-0.4 +/- 2.3 mV, P = 0.06). Thus, in the ARVC group, R-wave amplitude at year 5 was significantly lower than that in the control group (5.7 +/- 4.8 mV versus 12.5 +/- 4.5 mV, P = 0.001). In ARVC patients with ICDs, ventricular sensing is likely to deteriorate during long-term follow-up; however, in HCM patients, sensing may not deteriorate.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    1
    Citations
    NaN
    KQI
    []