Fat-Water Interface on Susceptibility-Weighted Imaging and Gradient-Echo Imaging: Comparison of Phantoms to Intracranial Lipomas

2013 
OBJECTIVE. In a clinical setting, lipoma can sometime show low signal intensity on susceptibility-weighted imaging (SWI) mimicking hemorrhage. The purpose of this study was to evaluate the fat-water interface chemical-shift artifacts between SWI and T2*-weighted imaging with a phantom study and evaluate SWI in lipoma cases. MATERIALS AND METHODS. SWI, magnitude, high-pass filtered phase, and T2*-weighted imaging of a lard-water phantom were evaluated in the in-phase, out-of phase, and standard partially out-of-phase TE settings used for clinical 3-T SWI (19.7, 20.9, and 20.0 ms, respectively) to identify the most prominent fat-water interface low signal. SWI of five cases of CNS lipoma were retrospectively evaluated by two neuroradiologists. RESULTS. TE at 19.7 ms (in-phase) showed the minimum fat-water interface low signal in the phase-encoding direction on magnitude, high-pass filtered phase, and SWI. TE at 20.9 ms (out-of-phase) showed the maximum fat-water interface in the phase-encoding direction on ...