A Test Chamber for Experimental Hydrogen Fluoride Exposure in Humans

1997 
An inhalation chamber was built to perform experimental studies with hydrogen fluoride (HF), other gases, and particulate matter. The present study sought to describe a new gas delivery system and the distribution and concentration of HF gas in the chamber. The aluminum chamber has a volume of 19.2 m3 and a variable ventilation rate of about 1 to 10 air changes per hour. The negative pressure difference between the chamber and outside air can be regulated from 0 to 300 Pa. HF was fed at concentrations of up to 4000 mg/m3 directly into the ventilation duct feeding the chamber through openings with diameters as small as 50 μm, oriented opposite to the airflow. Gas flow was varied from about 0.1 dm3/min at a pressure of 4 atm. The dilution factor of HF concentration from cylinder to chamber was on the order of 103 to 104. The standard deviation (SD) of the HF concentrations at a fixed measurement point during a 1-hour test was typically 0.05 mg/m3 at a time-weighted average (TWA) concentration of 2.66 mg/m3....