Development of a multidimensional gamma-spectrometer

2017 
A high-sensitivity multidimensional gamma-spectrometer is being developed within the shallow underground laboratory at Pacific Northwest National Laboratory (PNNL, USA). The system consists of two broad energy germanium detectors, inside a low-background shield, fitted with a cosmic veto system. The detector has advanced functionality, including operation in single or combined detector mode, with reductions in the cosmic background of 49.6% and Compton suppression of 6.5%. For selected radionuclides this provides increased peak identification, reductions in uncertainty of 27.6% and MDA improvements of 52.7%. The design uses commercially off-the-shelf components to provide a powerful solution for low-level nuclear measurements.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    28
    References
    7
    Citations
    NaN
    KQI
    []