Image distortion in MRI-based polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems

2002 
We have studied effects of MR (magnetic resonance) image distortion on polymergeldosimetry of Gamma Knife stereotactic radiosurgery systems. MRimages of BANG polymergel phantoms were acquired by using a Hahn spin-echo sequence and a fast 3D imaging GRASS sequence. Image artifacts were studied by varying the directions of frequency encoding and the receiver bandwidth. The phantoms were also CT(computed tomography) scanned. The studies showed that the measured dose distributions are shifted by 1.8±0.5 mm (2 pixels) in the frequency encoding direction. The magnitude of the shift is inversely proportional to the receiver bandwidth in agreement with theory. Comparison of MRI with CT showed the same image shift. We concluded that the discrepancy is caused by MRimage distortion due to a difference in susceptibility effects between the phantom and the fiducial markers of the Leksell localization box.